For the purpose of treating health checkups and recovery of patients in a super-aged society, hospitals use devices designed with a reduction circuit of electromagnetic waves associated with the specific absorption rate of electromagnetic waves absorbed by the human body. In this paper, we proposed a filter improvement design method capable of reducing electromagnetic waves. As a result of confirming the validity of the proposed technique through simulation and experimental results, the following result values were obtained. Applying the common-mode (CM) inductor 4 mH to a calibration circuit, noise decreased in a multiband spectrum. Using the differential mode(DM) inductor 40 μH element in the primary calibration circuit, the noise decreased by 15 dB or more in the 3 MHz band spectrum. Also, applying the Admittance Capacitance (Y-Cap) 10 nF element in the secondary calibration circuit resulted in the decrease by more than 30 dB in the band spectrum before 2 MHz. After using a common-mode inductor 4 mH element in the tertiary calibration circuit, it decreased by more than 15 dB in the band spectrum after 2 MHz.